RPT · 12 questions
- A technologist uses a portable ionization chamber with a movable beta shield to survey a work surface suspected of having mixed beta and gamma fields. With the shield closed, the meter reads 20 mR/hr; with the shield open, it reads 120 mR/hr. What does the difference between these two readings represent?
- Before carrying a portable Geiger-Mueller survey meter into a radiologically controlled area, a technician inspects the calibration sticker and confirms the date is current. What additional check must the technician perform immediately before relying on the meter?
- A technologist needs to survey a workbench for trace alpha contamination in an area with a moderate gamma background of 2 mR/hr. Which detector is best suited to detect the alpha emitters while rejecting the gamma background?
- A technician enters an area near an exposed high-activity gamma source with a standard Geiger-Mueller (GM) survey meter, and the analog needle suddenly drops toward zero despite a visibly unshielded source. What instrument limitation explains this behavior?
- A technologist uses a vented (unsealed) air-filled ionization chamber calibrated at sea level to measure gamma exposure rates at a high-altitude facility. Why must the meter readings be corrected for ambient temperature and barometric pressure?
- A technologist needs to measure accurate photon exposure rates (mR/hr) around a mixed field containing both 60 keV scatter and 1.25 MeV gamma rays. Why is a portable air-equivalent ionization chamber preferred over a standard Geiger-Mueller (GM) survey meter for this task?
- A radiation protection technologist needs to measure fast neutrons around a nuclear reactor shield using a Helium-3 (He-3) proportional counter. Why must the He-3 tube be surrounded by a thick polyethylene sphere or cylinder?
- While checking a handheld ZnS(Ag) alpha scintillation survey meter in a brightly lit room before entering a work area, a technician notices a tiny pinhole in the aluminized Mylar entrance window. What effect will this defect have on the instrument?
- When counting a high-activity beta-gamma source with a Geiger-Mueller (GM) survey meter, the observed count rate begins to fall noticeably below the true rate of ionizing events entering the tube. What instrument phenomenon causes this under-response?
- A health physicist needs to measure the average exposure rate outside a shielding wall around a pulsed medical linear accelerator that delivers radiation in brief microsecond bursts. Which portable survey instrument is best suited for this measurement?
- A health physics technician uses a handheld pancake Geiger-Mueller (GM) probe for routine personnel frisking at a Contamination Area exit. Why is this probe constructed with an ultra-thin mica entrance window rather than a standard metal wall?
- A technician is scanning a large equipment laydown area with a portable count-rate survey meter that has a fast/slow response-time switch. Which setting should the technician select while actively sweeping the probe to avoid missing small, localized hot spots?