- A scaler records 10,000 total counts during a 10-minute measurement of a long-lived sample. Assuming Poisson counting statistics, what is the one-sigma standard deviation of the total count?
- A technician counts a smear sample for 10 minutes and records 1,600 gross counts, then counts a clean background planchet for 5 minutes and records 225 counts. Assuming Poisson statistics, what are the net count rate and its one-sigma standard deviation?
- A technician collects ten consecutive 1-minute counts of a long-lived check source on a laboratory scaler and runs a chi-square test. The resulting chi-square value falls far above the expected upper critical limit. What does this indicate?
- A counting laboratory needs to cut the Minimum Detectable Activity (MDA) of a low-level smear counter in half using the same detector, geometry, and sample. Assuming equal sample and background counting times and a dominant background rate, how must the counting times be adjusted?
- A laboratory technician counts a long-lived check source for 4 minutes and records 1,600 total counts, giving a count rate of 400 cpm with a one-sigma standard deviation of 10 cpm. Assuming negligible background, how long must the technician count the same source to reduce the count rate's one-sigma standard deviation to 5 cpm?
- A sample is counted for 10 minutes and yields 1,600 gross counts (160 cpm). A 10-minute background count yields 900 counts (90 cpm). Assuming Poisson statistics, what is the net count rate and its one-sigma standard deviation?
- A non-paralyzable Geiger-Mueller counting system has a resolving dead time of 200 microseconds and records an observed count rate of 1,000 counts per second. What is the dead-time-corrected true count rate?
- In low-level radioactivity counting, why is the Lower Limit of Detection (LLD) set higher than the Critical Level (Lc) for the same counting system and time?
- A radiochemistry technician needs to count a low-background calibration check source long enough to achieve a one-sigma relative standard deviation (percent uncertainty) of 2.0 percent on the gross counts, assuming negligible background. What minimum number of total counts must the scaler accumulate?
- A technician counts a NIST-traceable Cesium-137 check source certified at 20,000 dpm for 5.0 minutes and records 27,000 gross counts. A 5.0-minute background count yields 2,000 counts. What is the detector's 4-pi counting efficiency for this geometry?
- When establishing the Critical Level (Lc) for a radioactivity counting system at a 5 percent false-positive probability, what does a net sample count exceeding Lc indicate?
- A technician has a fixed total counting time of 30 minutes to measure a sample with an approximate gross count rate of 400 cpm and a background count rate of 100 cpm. To minimize the statistical uncertainty of the net count rate, how should the 30 minutes be divided between the sample and background counts?