Quiz 14 Question 12 of 20

Morning PET/CT quality control passes on this unit: energy peaking, uniformity flood, and the CT air-water calibration are all within tolerance. A repeat study of a stable patient still shows SUVs consistently higher than baseline despite identical injected activity, uptake time, and decay correction. Which calibration check should the technologist run next to explain the shift?

Select an answer to reveal the explanation.

Motivation